-

A63045-UN-13AUG08
Seed Records

A63027-UN-13AUG08
Previous Record

A63026-UN-13AUG08
Next Record

A63812-UN-13NOV08
Left-Right Configuration Shown
A - Light Tab, View Record
B - Dark Tab, Hidden Record
C - Seed Selection Drop-down
D - Seed Type, Normal or Fine
E - Desired Rate
F - Target Travel Speed
G - Right Indicator
H - Calibration Result
I - Meter Gear Recommendation
J - Meter Value Recommendation
Select Menu button >> select Drill Application icon >> select Function Toggle until icon set 2 is displayed >> select Seed Records icon.
-
NOTE: The seed record page is different from each hopper configuration.
With a left-right tank configuration, the hoppers share one record
page. Other configurations have separate seed record pages per hopper.
Select Next Page and Previous Page to move between hoppers and records.
Multiple application records can be stored. A record with a
light-colored tab (A) is viewed. Select Next Record and Previous Record icons to view records
with dark-colored tabs (B). The tab with a flashing number is the
active record currently controlling the meter.
-
Select drop-down menu (C) and choose a seed.
-
IMPORTANT: Avoid machine damage and incorrect calibrations.
Follow screen commands. Electronic calibration procedure is different
from mechanical calibration. Calibration labels on machine are only
for mechanical calibration procedure.
Either Normal Seed or Fine Seed appears in location (D). Set
butterfly valve and changeover switch, on the blower housing, to “N”
for normal seed or “F” for fine seed.
-
Select drop-down box (E) and enter desired rate of seed delivery
in units displayed next to box.
-
Select drop-down box (F) and enter target travel speed in units
displayed next to box.
-
Select drop-down box (G) and enter fraction of area to simulate
for calibration. Do not exceed 1/10. The larger the area simulated
the more accurate the calibration, but the sample is larger and can
take longer to collect.
-
Indicator (H) displays an “X” until calibration
is completed properly. A checkmark appears after a successful calibration
and calibration test.
-
Set the meter gear according to the meter recommendation (I).
Red indicates to disengage micrometer. Green indicates to engage micrometer.
-
Set the meter rate according to the meter recommendation (J).
-
Place a collection container on a scale and zero scale reading.
-

ZX025540-UN-28MAY01

A63045-UN-13AUG08
Start Calibration

A63803-UN-11NOV08
Start/Accept Icon
A - Manifold
B - Wing Nut
Remove wing nut (B) and manifold (A) under meter selected
for calibration. Place a collection container under meter.
-
CAUTION: Avoid injury. Stand away from meters during calibration
and calibration test. Meters can engage unexpectedly.
Select Start Calibration icon. A screen
appears to indicate progress of meter filling with product.
-
Select Start icon to begin filling
meter. Wait for filling to complete or select “ESC” icon
to cancel.
-
Empty collection container and place container beneath meter.
-
The calibration screen appears. Select Start icon to begin calibration.
-
When the calibration progress bar is complete and meter stops,
weigh the sample collected.
-
Use the target rate and fraction of calibration area to calculate
the expected sample weight. If the sample weighs much different from
the expected amount, perform calibration again. If sample is close
to expected weight, enter sample weight in input box.
-
Select Accept icon.
-

A63049-UN-13AUG08
Calibration Test

A63803-UN-11NOV08
Start/Accept Icon
Select Calibration Test icon.
-
NOTE: Acceptable calibration area values range between 0.1 and
10.
Enter a calibration area to test in the input box and place
an empty collection container under meter.
-
Select Start icon.
-
When meter stops, weigh sample collected. Compare the amount
to the fraction of target rate entered into calibration area. If values
differ, perform another calibration. If values match, accept and confirm
data. Repeat test for additional meters.
Perform calibration on each meter for each product. If the same
product is used for different rate records, the identical collection
sample can be entered without performing additional calibrations.
Different products always require a calibration and calibration test.
If the same product is used in a tank but the quality is different,
a calibration test is recommended. A calibration is not required.