Calibrate the Meter Manually

ZX013873-UN-02DEC97
Seed Meter

ZX013874-UN-02DEC97
Blower
A - Housing
B - Meter Scale
C - Red Changeover Switch (N=Normal Seed; F=Fine Seed)
D - Spindle
E - Hopper Trapdoor
F - Rotary Slide Valve
G - Manifold
H - Spring Clip on Butterfly Valve (N=Normal Seed;
F=Fine Seed)
IMPORTANT: Avoid machine damage and incorrect calibrations.
If meter is controlled through an ISOBUS monitor, see Seed Calibration
in the ISOBUS Seed Monitor System section to calibrate the meter.
To calibrate the meter, refer to the illustrations, the calibration
procedure, and the calibration chart. Engage the micrometer system
to sow fine seed or small quantities of seed (See Engage Standard
Meter System or Micrometer in this section.).
IMPORTANT: Calibrate only when the drill is stationary. Switch
OFF the tractor and blower before calibration.
Calibration Procedure
Calibration procedure is as follows:
-
Determine seed type (normal seed or fine seed).
-
Set meter for normal seed or fine seed.
-
Refer to calibration chart for meter setting that corresponds
to desired seed rate.
-
The meter scale (B) can be set to any value from 0 to 110 for
normal seed and from 0 to 25 for fine seed. Turn the spindle (D) until
the desired setting is visible at the edge of housing (A).
-
IMPORTANT: On 6 m version, each seed meter must be set and calibrated.

ZX013873-UN-02DEC97
Seed Meter

ZX013874-UN-02DEC97
Blower

ZX013876-UN-02DEC97

ZX013875-UN-02DEC97
Manual Calibration with Handle
A - Housing
B - Metering Scale
C - Red Changeover Switch (N=Normal Seed; F=Fine Seed)
D - Spindle
E - Hopper Trapdoor
F - Rotary Slide Valve
G - Manifold
H - Spring Clip on Butterfly Valve (N=Normal Seed;
F=Fine Seed)
J - Wing Nut
K - Calibration Handle
Close the hopper trapdoor (E).
-
Fill the seed hopper with seed.
-
Remove wing nut (J) and manifold (G) from underneath the venturi
cone.
-
Place a collection bin underneath the venturi cone. This collects
a seed sample during calibration.
-
Remove the spring pin and pull off the drive shaft. Attach the
calibration handle (K) to the drive shaft as shown.
-
Turn the calibration handle 85 times counterclockwise at approximately
1 revolution per second. The sample of seed collected in the collection
bin corresponds to 1/10 hectares.
-
Weigh the seed sample collected.
-
Multiply weight in collection bin by 10 and compare to weight
shown in calibration chart for this seed and meter setting. If numbers
are not equal, calculate the percentage to increase or decrease meter
setting (see example from now on).
-
Repeat calibration procedure and weigh new sample until result
is acceptable.
-
Replace manifold (G) and wing nut (J).
-
Remove calibration handle, reconnect the drive shaft, and secure
it with the spring pin.
The meter has now been calibrated.
Example:
|
|
ZX,OM750A010218-19-20160505
|
|