Change The Meter Strip (Agitation or Smooth)
-

A83322-UN-19AUG14
A - Agitator Strip
B - Smooth Strip
Determine which agitator strip is need for the crop being
planted. The agitator strip (A) has a rough surface. The smooth strip
(B) has a smooth surface. (See ExactEmerge Meter Settings for
the recommended agitator strip by crop.)
-
Remove the meter from the row unit. (See Remove and Install
The ExactEmerge™ Seed Meter in the ExactEmerge™ Meter
Operation section.)
-
Open the meter dome. (See Clean Out The Vacuum Meter in the
ExactEmerge™ Meter Operation section.)
-
Remove the seed bowl. (See Clean Out The Vacuum Meter in the
ExactEmerge™ Meter Operation section.)
-

A83325-UN-19AUG14
A - Tabs
To release the agitator strip, squeeze the tabs (A) on the
back of the meter then pull out on the strip.
-

A83323-UN-19AUG14

A83324-UN-19AUG14

A83326-UN-19AUG14
A - Agitator Strip
B - Tabs
C - Holes
D - Tabs
To install the correct agitator strip (A) in the meter, align
the tabs (B) with the holes (C). Push the strip into position until
the tabs (D) latch into the holes.
-
Install the correct seed bowl. (See Install Seed Bowl in ExactEmerge™
Meter Operation section.)
-
NOTE: The meter hub height must be checked when the seed bowl
is changed.
Adjust the meter hub height. (See Adjust The Meter Hub Height
in the ExactEmerge™ Meter Operation section.)
-
Close the meter dome.
-
Install the meter on the row unit. (See Remove and Install The
ExactEmerge™ Seed Meter in the ExactEmerge™ Meter Operation
section.)
|
|
WP29706,00007E0-19-20160302
|
|