Change Moisture, Sample Rate and Yield Units Settings

NOTE: Page appears if equipped with moisture sensor.

Moisture Sensor Settings

From setup pages icon (A), rotate selection dial (B) until ON box (C) or OFF box (D) is highlighted:

  • Moisture Alarm ON Box - When checked, moisture alarm will sound when value drops below minimum setting (operator preset) and above maximum setting (determined by operator).
  • Moisture Alarm OFF Box - When checked, moisture alarm will not sound when value drops below minimum setting (operator preset) and above maximum setting (determined by operator).

When desired box is highlighted, press confirm switch (E).

Rotate selection dial until minimum box (F) or maximum box (G) is highlighted:

NOTE: This is used to determine setpoints (minimum and maximum) for activation of moisture alarm.

  • Minimum Moisture Alarm Box - Operator selects a moisture percentage for alarm to sound when value drops below preset value.
  • Maximum Moisture Alarm Box - Operator selects a moisture percentage for alarm to sound when value is above preset value.

When desired box is highlighted, press confirm switch.

Rotate selection dial until desired value is shown in box. Press confirm switch to save values.

H90869
H86934


A-Setup Pages Icon
B-Selection Dial
C-ON Box
D-OFF Box
E-Confirm Switch
F-Minimum Box
G-Maximum Box

OUO6075,0000049 -19-25JUN08-1/4


NOTE: Sample rate is how often moisture sensor takes crop samples. Some crops are abrasive and cycle times need to be lengthened (example: Rice).

Settings are stored in memory when key is cycled. If sample rate setting is set to zero (moisture sensor disabled) this setting is not saved and previous setting is restored after key is cycled.

Sample Rate Settings

Rotate selection dial until sample rate box (A) is highlighted, press confirm switch.

Rotate selection dial until desired value is shown in box. Press confirm switch to save values.

H90880


A-Sample Rate Box

OUO6075,0000049 -19-25JUN08-2/4


Moisture Correction/Fixed Moisture Settings

Rotate selection dial (A) until moisture correction box (B) or fixed moisture box (C) is highlighted:

IMPORTANT: Changing moisture correction will not change data already saved. After changes are made, all harvest information collected from that point will reflect the changes.

  • Moisture Correction Box - Set moisture correction to match readings from a customer or elevator certified moisture sensor.
  • Fixed Moisture Box - Moisture sensor reading will be disabled and forces preset moisture values.

When desired box is highlighted, press confirm switch (D).

Rotate selection dial until value/change box (E) is highlighted and press confirm switch.

Rotate selection dial until desired value is shown in box. Press confirm switch to save value.

H87194
H90871


A-Selection Dial
B-Moisture Correction Box
C-Fixed Moisture Box
D-Confirm Switch
E-Value/Change Box

OUO6075,0000049 -19-25JUN08-3/4


Yield Units Settings

Rotate selection dial (A) until yield units menu (B) is highlighted. Press confirm switch (C).

Rotate selection dial to choose desired yield units:

Bushels Pounds
Barrels Kilograms
Sacks Metric Tons
Hundred Weight Tons

 

Once desired yield units are chosen, press confirm switch.

NOTE: Operator selects desired global units from regional display settings page for desired area units, refer to Regional Settings, Time/Date Settings, Units of Measurement later in this section.


A-Selection Dial
B-Yield Units Menu
C-Confirm Switch

H87088
H90872

OUO6075,0000049 -19-25JUN08-4/4