Calibrate Solution Pump
The SprayStar system will automatically calibrate itself in
order to provide an optimal control for a particular situation.
The only time calibration is needed is when the A and B values
(represented by the digits following the A and B letters on display)
do not lie within their ranges or when one of the following has been
replaced:
-
Solution pump/hydraulic motor assembly
-
Proportional valve
-
Hydraulic pump
-
SRC
NOTE: The A and B ranges are:
High Flow
|
A = 2000-4000
|
350 - 850
|
|
B = 6000-8000
|
1300 - 1800
|
NOTE: If minimum B range value is not attained maximum solution
pump output is not available. See Troubleshooting section in this
manual.
Standard Flow
|
A = 2000-4000
|
350 - 850
|
|
B = 6500-9000
|
1400 - 2000
|
To auto calibrate the sprayer:
NOTE: Solution pump switch should be turned off to perform this
procedure. Hydraulic oil should be warm during the pump
calibration procedure.
-

N89545-UN-23SEP10
A - Calibration Softkey
B - Spray Calibration Tab
C - Calibration Icon
Operate engine at full throttle and warm hydraulic fluid
temperature to normal operating range of 60-88° C (140 -190°
F).
-
Fill solution tank with enough water or solution to keep pump
from running dry.
-
Select the Calibration softkey (A) to view the setup menu.
-
Select the Spray Calibration tab (B).
-
Select calibration icon (C) next to Pump Calibration. Pump Calibration
will appear.
-
NOTE: Warm hydraulic fluid temperature to normal operating range
before calibration and fill solution tank with enough water or solution
to keep pump from running dry.

N74766-UN-18JAN07
A - Enter Button
B - Enter Button
Follow instructions on first screen. When finished select
“Enter” button (A) to proceed to next screen.
-
NOTE: Pump calibration will take approximately 1-2 minutes.
The solution pump will turn off when auto calibration is done.
“Solution pump calibration is complete.” will be
displayed. Press “Enter” button (B) to return to Spray
Calibration tab.
|
|
OUO6092,0000439-19-20100922
|
|