Calibrate Pressure Sensor

N74946
N74946-UN-01FEB07

A - Pressure Reading


If pressure sensor (A) reading does not return to zero when there is no pressure or fluid in boom section shut-off valves, calibration of pressure sensor is necessary.

  1. N87485
    N87485-UN-25MAR10

    A - Calibration Softkey

    B - Spray Calibration Tab

    C - Calibration Icon

    Select the Calibration softkey (A) to view the setup menu.



  2. Select the Spray Calibration tab (B).

  3. Select calibration icon (C) next to Pressure Sensor Calibration. Pressure Sensor Calibration will appear.

  4. NOTE: Allow approximately five minutes for pressure in solution pump to dissipate before removing flowmeter.

    CAUTION: Do not drain solution onto the ground. Drain into a container. Drain solution in an area where people, animals, vegetation and water supply, etc., cannot be contaminated.

    N88060
    N88060-UN-06APR10

    A - Flowmeter Insert

    Slowly remove flowmeter insert(s). Catch fluid in a suitable container.



  5. N74765
    N74765-UN-18JAN07

    A - Button

    B - Enter Button

    C - Enter Button

    Follow instructions on first screen. When finished select the button (A) to proceed to next screen.



  6. Follow instructions on second screen. When finished select the enter button (B) to proceed to next screen.

  7. “Pressure sensor calibration is complete.” will be displayed. Press enter button (C) to return to Spray Calibration tab.

  8. Reinstall insert in flowmeter.

SW97031,00000C0-19-20110419