|
Enter Flowmeter Calibration number as stamped on valve.
-
Most flowmeters have an attached tag indicating the recommended calibration number. Enter this number as the initial flowmeter calibration value.
NOTE:
If the tag is no longer attached, start with a low calibration number such as 60, and change the value accordingly from that point.
NOTE:
For RAVEN flowmeters, divide the number on the tag by 10 before entering. E.g. "2650 L" = 265 ("700 gal" = 70).
NOTE:
The number on the RAVEN tag is pulses per 10 units of fluid and the value entered needs to be pulses per unit of fluid.
|
Choose unit of measurement from flowmeter units drop-down.
Check the pressure sensor box if sensor is installed. Pressure indicator will be displayed on main menu instead of flow rate when pressure sensor is checked.
To calibrate pressure sensor, press button and follow on screen instructions.
See CONTROLLER DIAGNOSTICS section for system energize process.
|