Row Sensor Calibration Procedure

This procedure is performed when the system is installed or after system is repaired. Row sensors must be installed and positioned against the at-rest stops.

1. Prepare for Calibration

Verify row sensors are installed with springs holding them in the at-rest position. Raise header to ensure that the row sensors are not contacting the ground. Combine must not be moving.


A-Steer Sensitivity (50-200)
B-AutoTrac Calibration
C-Row Guidance Option Installed
D-Row Guidance Sensor Calibration
E-Row Guidance Offset (50-150)
F-
G-Return

PC10466

AutoTrac Setup

OUO6050,0000F2F -19-27APR08-1/2


2. Press SETUP button on Original GreenStar Monitor. Press letter button pointed to by AutoTrac.

3. Calibrate Sensor At-Rest Voltages

Press Row Guidance Sensor Calibration button (E) to store the sensor at-rest voltages into SSU memory.

NOTE: GSD screen will now display the left sensor and right sensor voltages. The right sensor at-rest voltage should be greater than 2.5 volts. The left sensor at-rest voltage should be less than 2.5 volts.

4. End of Calibration

Exit row sensor calibration mode.


A-
B-
C-Left Sensor Cal (V) 1.17
D-Right Sensor Cal (V) 3.83
E-Calibrate Row Guidance Sensor
F-
G-Return

PC10467

Calibration Setup

OUO6050,0000F2F -19-27APR08-2/2