-

PC12277-UN-16SEP09
Control Valve Type

PC10645-UN-11OCT07
Dual Valve System
A - Control Valve Type Drop-Down Menu
Select Control Valve type from the drop-down menu. See the
following valve type descriptions.
NOTE: Valve type must be selected for system to work.
Select STANDARD or FAST for valve configurations with a control valve
and separate downstream on/off valve(s). Select FAST CLOSE for valve
configurations with one valve that combines control and on/off. Selecting
the wrong valve type may result in unexpected behavior and degraded
performance.
-

PC12985-UN-28OCT10
Control Valve Calibration
A - Control Valve Calibration
B - Tank Capacity
C - Tank Capacity Displayed
D - Tank Capacity Units
Enter Control Valve Calibration number into control valve
calibration input box.
Control Valve Calibration Table
|
Standard Valve Type
|
Valve Calibration Number (XXYZ)
|
|
RAVEN 165
|
2513
|
|
RAVEN 894
|
2513
|
|
RAVEN 125
|
2513
|
|
TEEJET 344B
|
1003
|
|
HARDI
|
7051
|
| |
|
Fast Valve Type
|
Valve Calibration Number (XXYZ)
|
|
RAVEN 177
|
0753
|
|
HINIKER Servo Valve (8160 Monitor Compatible)
|
0433
|
| |
|
Fast-Close Valve Type
|
Valve Calibration Number (XXYZ)
|
|
RAVEN 177
|
0753
|
|
HINIKER Servo Valve (8160 Monitor Compatible)
|
0433
|
NOTE: On a RAVEN valve, the last three numbers of the model
number are used to identify the valve type (e.g. RAVEN XXX894).
Use the recommended values in the Control Valve Calibration
table to initially set the control valve number. Evaluate machine
behavior during operation, and if necessary, adjust the value to attain
optimum system performance.
The Control Valve Calibration number is a 4 digit number used
to fine tune control characteristics. The number is in the form XXYZ,
and is defined as follows:
-
XX - Valve Speed.
-
The higher this value, the quicker the valve responds. If it
is set too high, the actual flow rate will oscillate continuously
around the target rate. If it is set too low, the target rate will
never be reached.
-
Y - Output Deadband.
-
Sets the minimum speed that the valve will ramp down toward
before stopping. The higher this value, the sooner the valve will
become stationary.
-
Z - Control Deadband:
-
Sets how close the actual flow rate must be to the target rate
to be considered acceptable. When the actual and target rates are
within this range, the valve will remain stationary. A higher value
allows for a larger disparity between actual and target rate.
Example of tuning the control valve for optimal
response:After
entering the initial starting value for the control valve (e.g. 2513)
and run a configuration test, the variance runs high (at or near 100%)
and the valve response seems sluggish. The next step in tuning may
be to increase the valve speed number (first 2 digits) from 25 to
35 and also increase the output deadband number (third digit) from
1 to 2 and retry the configuration test.
-
Enter maximum volume of tank in input box labeled tank capacity.
-
Select tank capacity displayed units from drop-down menu
-
Select tank capacity units from drop down menu.
-

PC12986-UN-28OCT10
Initial System Settings
A - Flowmeter Calibration Input Box
B - Flowmeter Units Drop-Down Menu
C - Calibrate Flowmeter Button
D - Pressure Sensor 1 Enable/Disable
E - Pressure Sensor 2 Enable/Disable
F - Calibrate Pressure Sensor Button
Enter Flowmeter Calibration number as stamped on valve.
-
Most flowmeters have an attached tag indicating the recommended
calibration number. Enter this number as the initial flowmeter calibration
value. It is not necessary to enter an additional product mass conversion
value as the controller does it automatically.
NOTE: If the tag is no longer attached, start with a low calibration
number such as 60, and change the value accordingly from that point.
NOTE: The number on the RAVEN tag is pulses per 10 units of
fluid. Ensure the 10 gal / 10 L unit is used.
-
Choose unit of measurement from flowmeter units drop-down.
-
Check the pressure sensor box if sensor is installed. If using
more than one pressure sensor, check the second box as well. Pressure
indicator will be displayed on main menu instead of flow rate when
pressure sensor is checked.
-
To calibrate pressure sensor, select button and follow on screen
instructions. See TEST section for details.