
ZX1043378-UN-21NOV09
Sensor Calibration - Step 1/2

ZX1043379-UN-21NOV09
Sensor Calibration - Step 2/2

ZX1043381-UN-17SEP09

ZX1043382-UN-17SEP09
A - Sensor Setup - 1/2
B - Contract Button
C - Contract Position Value (V)
D - Invert Steering Check Box
E - Next Button
F - Sensor Setup - 2/2
G - Broaden Button
H - Broaden Position Value (V)
I - Accept Button
J - Alarm Message
K - Alarm Message
L - Cancel Button
Press the Calibrate button on the RUN page.
Sensor setup 1/2 (A) is displayed with the following message:
”Contract the plough to its most narrow position using the button below. If the plough broadens, press the invert steering checkbox. Then press the next button.”
Press and hold Contract button (B) until the desired voltage level (C) is reached. The limit that can be set by the operator is 0.75 V or 4.25 V depending on the sensor mounting location.
NOTE: If the plough broadens instead of contracts, check the Invert Steering box (D).
Press Next button (E) to continue sensor position calibration process.
Sensor setup 2/2 (F) is displayed with the following message:
”Broaden the plough to its widest position using the button below. Then press the accept button.”
Press and hold Broaden button (G) until the desired voltage level (H) is reached. The limit that can be set by the operator is 0.75 V or 4.25 V depending on the sensor mounting location.
NOTE: Alarm messages ”Sensor value too close to previous setting.” (J) or ”Sensor value outside valid range. Consider a different mounting position.” (K) may appear while pressing Contract or Broaden button (B, G). The difference between the voltage settings must be at least 1V. Follow message instructions then press Accept button (I).
Once calibration is finished, press Accept (I) to validate calibration.